Interpretation of common endocrine laboratory tests: technical pitfalls, their mechanisms and practical considerations

Abstract

Pitfalls in hormonal assays are commonly seen in clinical practice and may lead to erroneous clinical impressions and treatments. In this article, we address common laboratory pitfalls encountered during evaluation of patients with real or presumed endocrine disorders including high dose hook effect and falsely normal prolactin in cases of macroprolactinomas, macroprolactinemia and falsely elevated prolactin, macrothyrotropinemia and falsely elevated TSH, heterophile antibodies leading to false elevation of hormonal concentration, biotin interference with different hormonal assays, cross-reactivity of steroid hormones immunoassays, and others. We describe the mechanisms of such laboratory pitfalls, review clinical scenarios in which they might occur, and discuss the ways to resolve such conundrums. The aim of this article is to present a learning material for the endocrine trainees and practitioners.

Background

Accurate hormonal assays play a significant role in the practice of endocrinology. A few decades ago, development of the radioimmunoassay (RIA) was awarded the Nobel prize as a revolutionary tool for measuring peptide hormones [1]. This was followed by measurement of the non-immunogenic steroid hormones [1,2,3]. Despite advances in laboratory techniques in the last few decades, pitfalls in endocrine testing can commonly happen distorting the clinical picture.

Various laboratory methods are used to assess endocrine problems including immunoassays and more recently, mass spectrometry. Immunoassays remain the most commonly used method to evaluate hormonal disorders [4]. They can be mainly divided into two groups: competitive and noncompetitive immunoassays.

In a two-step competitive immunoassay, antibodies to human hormone are generated in an animal of a certain species (rabbit, goat, guinea pig, etc.), this polyclonal first antibody is added to a sample of a patient’s serum or plasma together with a known amount of a radiolabeled hormone of interest that competes for binding to the first antibody with endogenous hormone. After incubation, the bound fraction is precipitated by a second antibody that is generated against the immunoglobulin G (IgG) belonging to the species in which the first antibody was produced, the supernatant is discarded, and the radioactivity of a pellet containing bound both labeled and endogenous hormones is measured. Understandably, the more endogenous hormone is contained in the sample, the less labeled hormone will be bound (this is the essence of competition). Thus, the weaker the signal, the more of the endogenous hormone was present in the sample and vice versa.

In clinical laboratories today, two step radioimmunoassay as described above has largely been replaced by non-isotopic single step competitive immunoassay. The antibody is immobilized on a solid surface, and a competition is set up by adding patient sample and a known concentration of labeled analyte. Chemiluminescent labels have become the dominant method for these assays (Fig. 1).

Fig. 1
figure 1

Illustration of single step competitive immunoassay. a In single step competitive immunoassay, capture antibodies are anchored to a solid phase in the testing tube. A certain concentration of labeled analyte (pre-prepared labeled antigen) is added to the testing tube together with the blood sample containing the studied hormone. b Both the studied analyte (the hormone) and the labeled analyte compete for the binding sites of the capture antibodies. After discarding the supernatant with both unbound endogenous and labeled analytes, the remaining signal of the bound labeled analyte is measured. The higher the concentration of the hormone (illustrated in green), the less labeled analyte will be bound, and thus, the less signal will be measured. In other words, the weaker the signal, the more of the endogenous hormone is present in the sample and vice versa (i.e. the signal strength is inversely proportional to the hormone concentration)

Noncompetitive assays employ two antibodies, but the mechanism of the assay is different. One monoclonal antibody is firmly attached to a solid surface of the test tube (the wall, plastic or glass ball, magnetic particles, etc.). It binds with the hormone in question and is called capture antibody. Another monoclonal antibody, labeled by various techniques (radioactivity, luminophore, fluorescent tag, etc.) is generated to a different epitope of the hormone in question and is called the signal antibody. It binds to the free epitope of the hormone in question already bound to the capture antibody and results in a formation of a so-called “sandwich”, i.e. a complex consisting of a capture antibody, hormone and signal antibody that is firmly attached to a solid surface. When the supernatant is discarded and the tube is gently washed, only the sandwich remains within it and the signal generated by the signal antibody is measured (Fig. 2). Understandably, the stronger the signal, the more hormone was contained in the serum sample and vice versa.

Fig. 2
figure 2

Illustration of non-competitive “sandwich” immunoassay. a In non-competitive “sandwich” immunoassay, capture antibodies are anchored to a solid phase in the testing tube. Then, the sample containing the studied hormone as well as the signal antibodies are added. b The hormone binds to the capture antibody from one end and to the signal antibody from the other end, forming antibody-hormone-antibody “sandwich”. The unbound signal antibodies is then washed out and the signal from the remaining “sandwiches” is measured. The higher the concentration of the hormone (illustrated in green), the more “sandwiches” will be formed, and thus, the higher signal will be measured. In other words, the stronger the signal, the more of the endogenous hormone is present in the sample and vice versa (i.e. the signal is directly proportional to the hormone concentration)

Limitations of these assays can be caused by many factors including pre-analytical sampling conditions, specificity of reagents used in the analysis, the presence of interfering or cross-reacting substances, and the presence of antibodies against either the reagents or the analyte. They can provide erroneous results leading the physician to pursue a wrong diagnosis and administer wrong treatment. When ordering a test, many factors need to be taken into consideration, including clinical suspicion of a problem, co-administration of various medications and reliability of the requested test. In this review article, we will highlight some pitfalls of commonly ordered endocrine tests that are important to consider for proper interpretation of a laboratory report. All pitfalls described here were encountered in our clinical practice in the Pituitary and Neuroendocrine Center at the University of Michigan, either during de-novo evaluation of a patient or, more often, during re-evaluation of patients who are erroneously diagnosed due to phantoms in the immunoassay.

The high dose hook effect

This phenomenon happens with the use of immunoassays to measure a certain hormone to be measured (analyte), in particular when using the two-site monoclonal “sandwich” assay [5]. Both serum sample and the signal antibody are added to the test tube simultaneously. As explained earlier, one epitope of the hormone binds to the capture antibody, while the other epitope of the hormone binds to the signal antibody, forming an antibody-hormone-antibody “sandwich” firmly attached to the solid surface. After the liquid component is discarded, the solid surface attached “sandwiches” elicit a signal that is directly proportional to the hormone concentration in the sample [6]. However, when the hormone concentration is exceedingly high or the amount of antibodies put by the manufacturer in the kit is low, the hormone saturates both the capture and signal antibodies preventing formation of the “sandwich”. As a result, after the liquid component is discarded, there will be only limited numbers of “sandwiches” attached to the solid surface and the detected signal will indicate low or only mildly elevated analyte concentration (Fig. 3). The shape of the binding curve gave the name “hook effect” to the phenomenon; with gradually-increasing analyte concentrations in the sample, the binding curve goes up, but at some critical point exceeding the capacity of the assay components, it starts “hooking down”. Thus, in cases of prolactinoma, for example, with exceedingly high prolactin levels, the false report of only mildly elevated prolactin may force the physician to make an erroneous diagnosis of a non-functioning pituitary tumor and subject the patient to unnecessary surgery (with potential complications) instead of prescribing dopamine agonists [7].

Fig. 3
figure 3

Illustration of the high dose hook effect. The left panel illustrates the non-competitive “sandwich” immunoassay with normal (or elevated within the tolerance of the assay kit) hormone concentration (see Fig. 2). The right panel illustrates the mechanism of the hook effect with exceedingly high hormone concentration. a The sample that contains remarkably elevated hormone concentration is added to the test tube which contains both capture and signal antibodies. b The studied hormone overwhelmingly saturates both the capture and signal antibodies preventing the formation of the “sandwiches”. c After the washout phase, only a few “sandwiches” will be left producing a low signal

How to prevent a potential hook effect phenomenon from making the clinical picture confusing? The presence of large deposits of neoplastic tissue is a necessary clinical clue. Understandably, only extremely high concentrations of analyte can cause the hook effect to occur. Since the magnitude of analyte concentration in the serum is as a rule proportionate to the size of a secreting tumor, only pituitary macroprolactinomas (usually above 4 cm) and malignant and widely metastatic tumors are likely to present this problem. The published list of other tests susceptible to this hook effect includes beta human chorionic gonadotropin (B-HCG) in patients with choriocarcinoma, thyroglobulin in thyroid cancer, and prostate-specific antigen in patient with metastatic prostate cancer [8,9,10]. In all the above situations, glossed over “normal” or “modestly elevated” serum markers can miss the diagnosis and withhold the needed therapy. The hook effect has not been described in cases of somatotropinomas (acromegaly) or corticotropinomas (Cushing disease) likely because these tumors are rarely big enough to produce astronomical hormone concentrations. Manufacturers of laboratory kits are aware of this problem and current kits often contain high enough concentrations of antibodies. For example, a prolactin assay employed in our hospital laboratory (ADVIA Centaur XP chemiluminometric assay by Siemens Diagnostics) is protected against the hook effect up to serum prolactin concentrations of 60,000 ng/mL. However, in one of our patients with a 9 cm skull base tumor, serum prolactin was 280,000 ng/mL [6]. Thus, a physician’s vigilance is still needed. Any large (above 4 cm) pituitary mass suspected of being an adenoma needs to be evaluated for the presence of hook effect.

One way of establishing true hormone concentration is to perform a sample dilution 1:100 or even more prior to adding it to the assay tube and then multiplying the results by the dilution factor. Another way is to incubate the serum sample with the capture antibody first, discard the supernatant containing the non-bound analyte (hormone) and only then add the signal antibody. Unfortunately, this latter technique is rarely an option with United States Food and Drug Administration (FDA) approved automatic immunoassays.

Macroprolactinemia

Prolactin is a human peptide hormone that is synthesized in the anterior pituitary gland in its pre-hormonal form. The pro-hormone undergoes cleavage and gets converted to the monomeric prolactin that weighs ~ 23 kDa. It is the most abundant form in the serum and known to be biologically and immunologically active [11]. This monomeric form is one of three forms of prolactin based on different molecular sizes; the two other forms include dimeric prolactin with a molecular weight of ~ 50 kDa and macroprolactin with a molecular weight of more than 100 kDa [12]. Macroprolactin, in fact, is a large prolactin-IgG antibody complex and is known to be biologically inactive [12, 13]. In normal subjects, the proportions of monomeric, dimeric, and macroprolactin were reportedly 85.8 ± 2.3, 9.1 ± 0.9, and 5.1% ± 1.7%, respectively [14].

Macroprolactinemia occurs when the high molecular weight macroprolactin predominates in the serum [15]. Macroprolactin is coupled with IgG forming a complex that has low receptor affinity and thus is biologically inactive [16]. The pathogenesis of macroprolactinemia remains unclear, Hattori et al. suggested that post-translational modifications can induce immunogenicity and formation of anti-prolactin autoantibodies resulting in high deposits of macroprolactin [17].

Macroprolactin interferes with most immunoassays used for prolactin level measurement leading to falsely elevated prolactin level. At the same time, macroprolactin does not yield to negative hypothalamic feedback, and thus, a true hyperprolactinemia can be further increased [18]. This can lead to misdiagnosis, mismanagement, and unnecessary consumption of medical resources. Among other many causes of hyperprolactinemia, macroprolactinemia should be considered especially when the clinical scenario lacks the clinical and the radiological evidence of hyperprolactinemia. In some studies, the prevalence of macroprolactinemia was as high as 26% of patients with apparent hyperprolactinemia [19]. It is important to note that patients with systemic lupus erythematosus (SLE) are especially prone to developing macroprolactinemia due to the presence of anti-prolactin antibodies; an estimated one-third of patients with SLE and hyperprolactinemia were found to have macroprolactinemia [20].

The gel filtration chromatography (GFC) remains the gold standard procedure to differentiate between different molecular forms of prolactin. However, this method is time and labor intensive [21]. Polyethylene glycol (PEG) precipitation has become a more commonly used screening method, given its ease and cost effectiveness. Adding PEG precipitates macroprolactin, leaving monomeric prolactin in the supernatant. Macroprolactinemia is usually suspected when PEG-precipitable prolactin exceeds 60% of the total prolactin, in other words, when the monomeric prolactin in the supernatant is less than 40% [22].

Macrothyrotropinemia (macro-TSH)

Thyroid stimulating hormone (TSH) is a commonly measured hormone in clinical practice. Similar to macroprolactin, anti-TSH autoantibodies form an antigen-antibody complex consistent of TSH and anti-TSH, forming what is known as macro-TSH with low bioactivity. This complex may affect some commercially available TSH assays, resulting in falsely elevated TSH levels [23].

Patients are usually euthyroid with normal thyroxine (T4) and triiodothyronine (T3) levels. Macro-TSH may mimic primary hypothyroidism and can be very challenging to suspect. This phenomenon is less prevalent compared to macroprolactinemia; in a study done by Mills et al., the prevalence of elevated TSH due to macro-TSH was found to be 0.6% [24].

In contrast to macroprolactinemia, PEG-precipitation method was not found reliable; nonspecific precipitation ratios were much higher for TSH than prolactin [25]. This finding concluded that macro-TSH should be suspected when PEG-precipitable TSH exceeds 90%, especially, when serum TSH is greater than 10 mU/L. Confirmation can be conducted by using gel filtration chromatography.

Heterophile antibodies

Heterophile antibodies (HAB) are antibodies that are formed due to exposure to external antigens. Animal antigens can be involved in forming what is called human anti-animal antibodies. A common antibody that falls in this category is human anti-mouse antibodies (HAMA) [26]. They are endogenous antibodies that form against murine monoclonal immunoglobulin [27]. In immunometric “sandwich” assays, HAMA can form a bridge between the capture and signal antibodies (Fig. 4), forming more “sandwiches”, leading to a falsely elevated signal of the studied analyte [28].

Fig. 4
figure 4

Illustration of human anti-mouse antibodies (HAMA) interference with immunoassay. The left panel illustrates the non-competitive “sandwich” immunoassay without the presence of HAMA in the sample (see Fig. 2). The right panel illustrates the mechanism of the HAMA interference with immunoassay. a The sample containing the studied hormone and the HAMA is added to the test tube which contains both capture and signal antibodies. b In addition to the correct formation of “sandwiches” (capture antibody- hormone-signal antibody), the HAMA forms a bridge between the capture antibody and the signal antibody forming antibody-HAMA-antibody “sandwiches”. As a result, more signal will be measured, and thus, false elevation of the studied hormone will be reported

The prevalence of HAB wildly varied in different studies. In a study by Koshida et al., overall, the prevalence of HAMAs in randomly collected samples was 11.7% [29]. These antibodies can interfere with a myriad of immunoassays including thyroid stimulating hormone, human chorionic gonadotropin (hCG), a-fetoproteins (AFP), cancer antigen 125 (CA-125), creatine kinase-muscle and brain (CK-MB) isozyme, and troponin among others [29, 30]. It is noteworthy that rheumatoid factor is an endogenous antibody that can mimic HAB and interfere with such immunoassays [31, 32].

An unexpected elevation of a hormone raises the suspicion of such issue. Subclinical hypothyroidism is commonly seen in endocrinology practice; in the presence of HAMA, TSH can be falsely elevated. In such case, a hint for the presence of these antibodies is failure of TSH normalization despite escalating doses of levothyroxine to the thyrotoxic range [33]. Patients who interact with animals, such as veterinarians or technicians studying mice, or just people living in mice-infested houses, can add a clue to the clinical picture.

When interference is suspected, several maneuvers can be undertaken to overcome such issue. Samples can often be retested with a different immunoassay or methodology since a number of commercially available assays has been developed to minimize the effects of HAMA [34]. Analyzing the sample with heterophile blocking reagent or diluting the sample and testing for linearity can also be done [35]. Measuring HAB (or HAMA) may identify patients who had exposure to a certain immunogen, however, this method has limited value and cannot prove HAB as the cause of an elevated result [30]. Communication with the laboratory performing the assay can be helpful in this situation.

Biotin interference with hormonal assays

Biotin, or vitamin B7, is a water-soluble vitamin that is involved in many enzymatic activities that regulate metabolism of fat, carbohydrates and amino acids [36]. Biotin is available in many plant and animal based sources of food. Despite the fact that biotin deficiency is rare, it is available in most of over the counter multivitamins preparations. Biotin supplementation has been increasingly used as an enhancer of skin, nails and hair health, despite the lack of evidence that supports that [37].

Biotin-streptavidin detection method is commonly used in many immunoassays [38]. Biotinylated antibodies (capture antibodies) bind strongly to streptavidin that anchors those antibodies to the solid phase of the assay. High biotin concentration in the serum interferes with this bond and alters the expected results [39]. In non-competitive “sandwich” immunoassays, the studied analyte (or hormone) is sandwiched between the capture and signal antibodies. In the presence of high biotin concentration in the specimen, biotin saturates streptavidin binding sites which alters the adherence between the capture (biotinylated) antibodies and streptavidin, leading to a low signal after the wash off phase (i.e. falsely low concentration of the analyte). In contrast, as in competitive assay, the studied analyte competes with the labeled analyte to bind to the specific antibodies. That means, the higher the concentration of the endogenous analyte, the less labeled hormone will bind to the antibody. In other words, the remaining signal after the wash off phase is inversely proportional to the analyte concentration. Thus, having high concentration of biotin in the specimen leads to lower signal of the labeled analyte, which maybe interpreted as falsely high concentration of the studied analyte in competitive immunoassays [40, 41].

The magnitude of biotin interference varies based on the concentration of biotin in the specimen, results may be distorted to be either falsely high or falsely low depending on the format of the assay used [42]. A broad spectrum of tests beyond hormonal assays can be affected by the presence of excess biotin in the serum, most commonly, thyroid function test; especially in the presence of a discrepancy between the TSH and free T4 levels. It is important to verify biotin use in patients, especially, if they present with non-matching clinical and biochemical picture. Biotin is excreted by the kidney with a half life between 8 and 16 h [43]. A simple way to overcome such interference is to stop biotin use for a few days before repeating the test. Another way is to use an alternative assay that does not depend on biotin-streptavidin technology. Not all immunoassay manufacturers use biotin-streptavidin coupling, so communication with the laboratory for guidance when interference is suspected is important. Some manufacturers incorporate reagents to eliminate the effect of exogenous biotin on the assay [44].

Cross-reactivity of steroid hormones immunoassays

Immunoassays are widely used to measure steroid hormones in different body fluids including serum, saliva and urine. A major pitfall of using immunoassays, especially competitive immunoassays, in measuring steroid hormones is cross-reactivity with other structurally similar molecules or compounds that can be either endogenous or exogenous in source [35]. This compromise can lead to inaccurate values of the measured hormone leading to altered clinical impression.

In certain clinical situations, such as congenital adrenal hyperplasia (CAH), adrenal steroid intermediaries, 11-deoxycortisol and 17-hydroxyprogesterone for example, may rise in the serum and interfere with cortisol immunoassays [45]. In a study by Monaghan el al, exogenous inhibition of steroidogenesis using metyrapone (inhibiting 11β-hydroxylase) in patients with Cushing’s syndrome resulted in false elevation of serum cortisol due to the accumulation of 11-deoxycortisol [46].

A situation that we have encountered several times in our clinic is a referral of a patient with the diagnosis of Cushing’s syndrome; the patient is treated with prednisone for some reason and often is, indeed, clinically cushingoid. Moreover, plasma and urinary cortisol levels are high (or, at least “not suppressed”). In combination with undetectable adrenocorticotropic hormone (ACTH) levels, that often leads to falsely suspected endogenous Cushing’s syndrome and almost always fruitless search for an adrenal tumor. The commonly overlooked pitfall is that most cortisol immunoassays have cross-reactivity to prednisolone, the active form of prednisone, ranging from modest to significant. Thus, prednisolone cross-reacts in the cortisol assay and “elevated cortisol” reported by the laboratory is in fact predictably-present prednisolone.

Other steroid hormones assays, such as testosterone and estradiol, are also subject to cross-reactivity if structurally similar compounds are present in the serum. When suspected, liquid chromatography-tandem mass spectrometry (LC-MS/MS) has became a well-known and more available technique providing more accurate measurement of the targeted steroid hormones to overcome such quandary [47].

Dexamethasone suppression test and cytochrome P450 3A4 (CYP3A4)

The overnight one milligram dexamethasone suppression test (DST) is commonly used to screen for Cushing’s syndrome [48]. Failure to suppress morning cortisol level below 1.8 mcg/dL raises such suspicion [49]. Serum dexamethasone has to reach certain level to be able to suppress the hypothalamic-pituitary-adrenal (HPA) axis, and thus, cortisol secretion. Certain conditions can impede cortisol suppression, including decreased dexamethasone absorption, enhanced dexamethasone metabolism and clearance, or in cases of pseudo-Cushing, such as psychiatric illness, obesity, alcoholism, and others [50, 51].

Dexamethasone is extensively metabolized via CYP3A4, in human gut and liver mainly into 6-hydroxydexamethasone and other metabolites [52]. Induction of CYP3A4 may result in fast dexamethasone metabolism, incomplete HPA axis suppression, and thus, false positive DST [53]. Various medications enhance CYP3A4 activity and lower serum dexamethasone levels such as phenobarbital, phenytoin, carbamazepine, primidone, mitotane, enzalutamide, apalutamide, rifampin, pioglitazone, St John’s wort and other medications [54,55,56,57,58,59]. It is noteworthy to mention that medications that increase cortisol binding globulin (CBG), like estrogen, may also lead to falsely positive DST by increasing the total cortisol level [60].

Before evaluation for possible Cushing’s via DST, it is essential to review patient’s comorbidities and medications that might lead to false positive test. It is also important to measure dexamethasone concentration in the serum at the time of measuring morning cortisol after suppression, this can provide an idea whether the desired dexamethasone concentration was actually achieved. If one milligram DST is doubtful, other screening tests can be pursued, such as twenty-four-hour urinary free cortisol (UFC), late-night salivary cortisol, late-night serum cortisol (if feasible), or two-day two milligram DST [50, 59].

Analog methods for free testosterone and thyroxine

These hormones are frequently measured in clinical practice. However, one must remember that significant proportions of their measured total concentrations circulate in bound forms, being attached to their specific binding proteins sex hormone binding globulin (SHBG) and thyroxine-binding globulin (TBG) respectively, and therefore are biologically inactive. Some proportions of them are weakly bound to albumin, and dissociate from it easily; thus they do exert biological effects. Any alterations in the concentrations of SHBG and/or TBG will falsely increase or decrease total concentrations of their corresponding hormones and non-specific binding of some drugs may also do the same. Equilibrium dialysis is the current gold standard for measuring free hormone concentrations, but is very time and effort consuming. For that reason, the so-called “analog methods” have been developed; they are based on the competition of radiolabeled analogs of the above hormones with (allegedly) their free fractions. In reality, however, free testosterone concentrations by analog methods are linked not to the true free, but to the total testosterone [61]. For that reason the Endocrine Society did not endorse the universal use of analog free testosterone assays in clinical practice [62]. The so-called “bio-available” testosterone assay is currently widely used in clinical laboratories and gives information on the sum of free and albumin bound testosterone. It correlates sufficiently well with free testosterone measurements by equilibrium dialysis and is a useful index of biological changes [62].

Free thyroxine measurement by analog methods is relatively robust and is widely used. However, it is still a subject to erroneous results by some platforms due to the presence of familial dysalbuminemic hyperthyroxinemia, anti-rutenium interference, thyroid hormone autoantibodies, etc. [63].

Unknown interfering substances

Occasionally, a patient may present with very confusing hormonal data that are often incompatible with the clinical picture and all the above mentioned reasons for that prove negative. In these infrequent cases, careful clinical examination, understanding of normal physiology of hormone synthesis, secretion and metabolism remain the only way to find an appropriate test(s) that suggest the presence of an unknown interfering substance [64]. Fortunately, such situations, although frustrating, are extremely rare.

Conclusions

Laboratory diagnosis of endocrine diseases is an indispensable tool in endocrine practice. It should always be done in conjunction with clinical assessment of the patient, including complaints, history of associated diseases, and concomitant medications as well as careful physical examination. Any discrepancy between clinical and laboratory data deserves careful attention. Blind reliance on laboratory reports often leads to erroneous diagnosis and treatment. Laboratory personnel have neither contact with the patient nor can they access the patient’s history, or drug therapy. Thus, it is the responsibility of the treating physician, especially the endocrine consult, to synthesize the entire information necessary for proper interpretation of the hormonal tests.

Unfortunately, many endocrine training programs do not include actual exposure of the trainees to the intricacies of laboratory diagnosis and do not require competency in the interpretation of hormonal tests. This needs to become a required part of training programs. This review is only a partial introduction to the most basic and commonly encountered potential pitfalls in laboratory practice. We believe that active involvement of clinical pathology mentors in the comprehensive training of future endocrinologists is essential.

Availability of data and materials

Not applicable.

Abbreviations

ACTH:

Adrenocorticotropic Hormone

B-HCG:

Beta Human Chorionic Gonadotropin bHCG

CAH:

Congenital Adrenal Hyperplasia

CBG:

Cortisol Binding Globulin

CYP3A4:

Cytochrome P450 3A4

DST:

Dexamethasone Suppression Test

FDA:

United States Food and Drug Administration

GFC:

Gel Filtration Chromatography

HAB:

Heterophile Antibodies

HAMA:

Human Anti-Mouse Antibodies

HPA:

Hypothalamic-Pituitary-Adrenal

IFMA:

Immunofluorometric Assay

ILMA:

Immunoluminometric Assay

IRMA:

Immunoradiometric Assay

LC-MS/MS:

Liquid Chromatography-tandem Mass Spectrometry

PEG:

Polyethylene Glycol

RIA:

Radioimmunoassay

SHBG:

Sex Hormone Binding Globulin

SLE:

Systemic Lupus Erythematosus

T3:

Triiodothyronine

T4:

Thyroxine

TBG:

Thyroxine Binding Globulin

TSH:

Thyroid Stimulating Hormone

UFC:

Urinary Free Cortisol

References

  1. Yalow RS, Berson SA. Assay of plasma insulin in human subjects by immunological methods. Nature. 1959;184(Suppl 21):1648–9.

    Article  CAS  Google Scholar 

  2. Abraham GE. Solid-phase radioimmunoassay of estradiol-17 beta. J Clin Endocrinol Metab. 1969;29(6):866–70.

    Article  CAS  Google Scholar 

  3. Handelsman DJ, Wartofsky L. Requirement for mass spectrometry sex steroid assays in the journal of clinical endocrinology and metabolism. J Clin Endocrinol Metab. 2013;98(10):3971–3. https://doi.org/10.1210/jc.2013-3375.

    Article  CAS  PubMed  Google Scholar 

  4. Taylor AE, Keevil B, Huhtaniemi IT. Mass spectrometry and immunoassay: how to measure steroid hormones today and tomorrow. Eur J Endocrinol. 2015;173(2):D1–12. https://doi.org/10.1530/EJE-15-0338.

    Article  CAS  PubMed  Google Scholar 

  5. Tate J, Ward G. Interferences in immunoassay. Clin Biochem Rev. 2004;25(2):105–20.

    PubMed  PubMed Central  Google Scholar 

  6. Barkan AL, Chandler WF. Giant pituitary prolactinoma with falsely low serum prolactin: the pitfall of the "high-dose hook effect": case report. Neurosurgery. 1998;42(4):913–5 discussion 915–6.

    Article  CAS  Google Scholar 

  7. Frieze TW, Mong DP, Koops MK. "hook effect" in prolactinomas: case report and review of literature. Endocr Pract. 2002;8(4):296–303.

    Article  Google Scholar 

  8. Levavi H, Neri A, Bar J, Regev D, Nordenberg J, Ovadia J. "Hook effect" in complete hydatidiform molar pregnancy: a falsely low level of beta-HCG. Obstet Gynecol. 1993;82(4 Pt 2 Suppl):720–1.

    CAS  PubMed  Google Scholar 

  9. Morgenthaler NG, Froehlich J, Rendl J, Willnich M, Alonso C, Bergmann A, Reiners C. Technical evaluation of a new immunoradiometric and a new immunoluminometric assay for thyroglobulin. Clin Chem. 2002;48(7):1077–83.

    CAS  PubMed  Google Scholar 

  10. Vaidya HC, Wolf BA, Garrett N, Catalona WJ, Clayman RV, Nahm MH. Extremely high values of prostate-specific antigen in patients with adenocarcinoma of the prostate; demonstration of the "hook effect". Clin Chem. 1988;34(10):2175–7.

    CAS  PubMed  Google Scholar 

  11. Vaishya R, Gupta R, Arora S. Macroprolactin; a frequent cause of misdiagnosed hyperprolactinemia in clinical practice. J Reprod Infertil. 2010;11(3):161–7.

    PubMed Central  Google Scholar 

  12. Sadideen H, Swaminathan R. Macroprolactin: what is it and what is its importance? Int J Clin Pract. 2006;60(4):457–61.

    Article  CAS  Google Scholar 

  13. Suh HK, Frantz AG. Size heterogeneity of human prolactin in plasma and pituitary extracts. J Clin Endocrinol Metab. 1974;39(5):928–35.

    Article  CAS  Google Scholar 

  14. Garnier PE, Aubert ML, Kaplan SL, Grumbach MM. Heterogeneity of pituitary and plasma prolactin in man: decreased affinity of "big" prolactin in a radioreceptor assay and evidence for its secretion. J Clin Endocrinol Metab. 1978;47(6):1273–81.

    Article  CAS  Google Scholar 

  15. Suliman AM, Smith TP, Gibney J, McKenna TJ. Frequent misdiagnosis and mismanagement of hyperprolactinemic patients before the introduction of macroprolactin screening: application of a new strict laboratory definition of macroprolactinemia. Clin Chem. 2003;49(9):1504–9.

    Article  CAS  Google Scholar 

  16. Hattori N, Ishihara T, Ikekubo K, Moridera K, Hino M, Kurahachi H. Autoantibody to human prolactin in patients with idiopathic hyperprolactinemia. J Clin Endocrinol Metab. 1992;75(5):1226–9.

    CAS  PubMed  Google Scholar 

  17. Hattori N, Ikekubo K, Nakaya Y, Kitagawa K, Inagaki C. Immunoglobulin G subclasses and prolactin (PRL) isoforms in macroprolactinemia due to anti-PRL autoantibodies. J Clin Endocrinol Metab. 2005;90(5):3036–44.

    Article  CAS  Google Scholar 

  18. Cavaco B, Prazeres S, Santos MA, Sobrinho LG, Leite V. Hyperprolactinemia due to big-big prolactin is differently detected by commercially available immunoassays. J Endocrinol Investig. 1999;22(3):203–8.

    Article  CAS  Google Scholar 

  19. Bjøro T, Mørkrid L, Wergeland R, Turter A, Kvistborg A, Sand T, Torjesen P. Frequency of hyperprolactinaemia due to large molecular weight prolactin (150-170 kD PRL). Scand J Clin Lab Invest. 1995;55(2):139–47.

    Article  Google Scholar 

  20. Leaños-Miranda A, Pascoe-Lira D, Chávez-Rueda KA, Blanco-Favela F. Detection of macroprolactinemia with the polyethylene glycol precipitation test in systemic lupus erythematosus patients with hyperprolactinemia. Lupus. 2001;10(5):340–5.

    Article  Google Scholar 

  21. Vallette-Kasic S, Morange-Ramos I, Selim A, Gunz G, Morange S, Enjalbert A, Martin PM, Jaquet P, Brue T. Macroprolactinemia revisited: a study on 106 patients. J Clin Endocrinol Metab. 2002;87(2):581–8.

    Article  CAS  Google Scholar 

  22. Beltran L, Fahie-Wilson MN, McKenna TJ, Kavanagh L, Smith TP. Serum total prolactin and monomeric prolactin reference intervals determined by precipitation with polyethylene glycol: evaluation and validation on common immunoassay platforms. Clin Chem. 2008;54(10):1673–81. https://doi.org/10.1373/clinchem.2008.105312.

    Article  CAS  PubMed  Google Scholar 

  23. Halsall DJ, Fahie-Wilson MN, Hall SK, Barker P, Anderson J, Gama R, Chatterjee VK. Macro thyrotropin -IgG complex causes factitious increases in thyroid-stimulating hormone screening tests in a neonate and mother. Clin Chem. 2006;52(10):1968–9 author reply 1969-70.

    Article  CAS  Google Scholar 

  24. Mills F, Jeffery J, Mackenzie P, Cranfield A, Ayling RM. An immunoglobulin G complexed form of thyroid-stimulating hormone (macro thyroid-stimulating hormone) is a cause of elevated serum thyroid-stimulating hormone concentration. Ann Clin Biochem. 2013;50(Pt 5):416–20. https://doi.org/10.1177/0004563213476271.

    Article  CAS  PubMed  Google Scholar 

  25. Hattori N, Ishihara T, Yamagami K, Shimatsu A. Macro TSH in patients with subclinical hypothyroidism. Clin Endocrinol. 2015;83(6):923–30. https://doi.org/10.1111/cen.12643.

    Article  CAS  Google Scholar 

  26. Kricka LJ. Human anti-animal antibody interferences in immunological assays. Clin Chem. 1999;45(7):942–56.

    CAS  PubMed  Google Scholar 

  27. Legouffe E, Liautard J, Gaillard JP, Rossi JF, Wijdenes J, Bataille R, Klein B, Brochier J. Human anti-mouse antibody response to the injection of murine monoclonal antibodies against IL-6. Clin Exp Immunol. 1994;98(2):323–9.

    Article  CAS  Google Scholar 

  28. Preissner CM, O'Kane DJ, Singh RJ, Morris JC, Grebe SK. Phantoms in the assay tube: heterophile antibody interferences in serum thyroglobulin assays. J Clin Endocrinol Metab. 2003;88(7):3069–74.

    Article  CAS  Google Scholar 

  29. Koshida S, Asanuma K, Kuribayashi K, Goto M, Tsuji N, Kobayashi D, Tanaka M, Watanabe N. Prevalence of human anti-mouse antibodies (HAMAs) in routine examinations. Clin Chim Acta. 2010;411(5–6):391–4. https://doi.org/10.1016/j.cca.2009.12.006.

    Article  CAS  PubMed  Google Scholar 

  30. Bolstad N, Warren DJ, Nustad K. Heterophilic antibody interference in immunometric assays. Best Pract Res Clin Endocrinol Metab. 2013;27(5):647–61. https://doi.org/10.1016/j.beem.2013.05.011.

    Article  CAS  PubMed  Google Scholar 

  31. Norden AG, Jackson RA, Norden LE, Griffin AJ, Barnes MA, Little JA. Misleading results from immunoassays of serum free thyroxine in the presence of rheumatoid factor. Clin Chem. 1997;43(6 Pt 1):957–62.

    CAS  PubMed  Google Scholar 

  32. Mongolu S, Armston AE, Mozley E, Nasruddin A. Heterophilic antibody interference affecting multiple hormone assays: is it due to rheumatoid factor? Scand J Clin Lab Invest. 2016;76(3):240–2. https://doi.org/10.3109/00365513.2016.1143113.

    Article  CAS  PubMed  Google Scholar 

  33. Santhana Krishnan SG, Pathalapati R, Kaplan L, Cobbs RK. Falsely raised TSH levels due to human anti-mouse antibody interfering with thyrotropin assay. Postgrad Med J. 2006;82(973):e27.

    Article  CAS  Google Scholar 

  34. Klee GG. Human anti-mouse antibodies. Arch Pathol Lab Med. 2000;124(6):921–3.

    CAS  PubMed  Google Scholar 

  35. Klee GG. Interferences in hormone immunoassays. Clin Lab Med. 2004;24(1):1–18.

    Article  Google Scholar 

  36. Mock DM, de Lorimer AA, Liebman WM, Sweetman L, Baker H. Biotin deficiency: an unusual complication of parenteral alimentation. N Engl J Med. 1981;304(14):820–3.

    Article  CAS  Google Scholar 

  37. Soleymani T, Lo Sicco K, Shapiro J. The infatuation with biotin supplementation: is there truth behind its rising popularity? A comparative analysis of clinical efficacy versus social popularity. J Drugs Dermatol. 2017;16(5):496–500.

    CAS  PubMed  Google Scholar 

  38. Diamandis EP, Christopoulos TK. The biotin-(strept) avidin system: principles and applications in biotechnology. Clin Chem. 1991;37(5):625–36.

    CAS  PubMed  Google Scholar 

  39. Colon PJ, Greene DN. Biotin interference in clinical immunoassays. The Journal of Applied Laboratory Medicine. 2018. https://doi.org/10.1373/jalm.2017.024257.

    Article  CAS  Google Scholar 

  40. Li D, Radulescu A, Shrestha RT, Root M, Karger AB, Killeen AA, Hodges JS, Fan SL, Ferguson A, Garg U, Sokoll LJ, Burmeister LA. Association of Biotin Ingestion with Performance of hormone and nonhormone assays in healthy adults. JAMA. 2017;318(12):1150–60. https://doi.org/10.1001/jama.2017.13705.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  41. Samarasinghe S, Meah F, Singh V, Basit A, Emanuele N, Emanuele MA, Mazhari A, Holmes EW. Biotin interference with routine clinical immunoassays: understand the causes and mitigate the risks. Endocr Pract. 2017;23(8):989–98. https://doi.org/10.4158/EP171761.RA.

    Article  PubMed  Google Scholar 

  42. Chun KY. Biotin interference in diagnostic tests. Clin Chem. 2017;63(2):619–20. https://doi.org/10.1373/clinchem.2016.267286.

    Article  CAS  PubMed  Google Scholar 

  43. Zempleni J, Mock DM. Biotin biochemistry and human requirements. J Nutr Biochem. 1999;10(3):128–38.

    Article  CAS  Google Scholar 

  44. Lam L, Kyle CV. A simple method to detect biotin interference on immunoassays. Clin Chem Lab Med. 2017;55(6):e104–6. https://doi.org/10.1515/cclm-2017-0059.

    Article  CAS  PubMed  Google Scholar 

  45. Roberts RF, Roberts WL. Performance characteristics of five automated serum cortisol immunoassays. Clin Biochem. 2004;37(6):489–93.

    Article  CAS  Google Scholar 

  46. Monaghan PJ, Owen LJ, Trainer PJ, Brabant G, Keevil BG, Darby D. Comparison of serum cortisol measurement by immunoassay and liquid chromatography-tandem mass spectrometry in patients receiving the 11β-hydroxylase inhibitor metyrapone. Ann Clin Biochem. 2011;48(Pt 5):441–6. https://doi.org/10.1258/acb.2011.011014.

    Article  CAS  PubMed  Google Scholar 

  47. Keevil BG. LC-MS/MS analysis of steroids in the clinical laboratory. Clin Biochem. 2016;49(13–14):989–97. https://doi.org/10.1016/j.clinbiochem.2016.04.009.

    Article  CAS  PubMed  Google Scholar 

  48. Arnaldi G, Angeli A, Atkinson AB, Bertagna X, Cavagnini F, Chrousos GP, Fava GA, Findling JW, Gaillard RC, Grossman AB, Kola B, Lacroix A, Mancini T, Mantero F, Newell-Price J, Nieman LK, Sonino N, Vance ML, Giustina A, Boscaro M. Diagnosis and complications of Cushing's syndrome: a consensus statement. J Clin Endocrinol Metab. 2003;88(12):5593–602.

    Article  CAS  Google Scholar 

  49. Newell-Price J, Trainer P, Besser M, Grossman A. The diagnosis and differential diagnosis of Cushing's syndrome and pseudo-Cushing's states. Endocr Rev. 1998;19(5):647–72.

    CAS  PubMed  Google Scholar 

  50. Boscaro M, Barzon L, Sonino N. The diagnosis of Cushing's syndrome: atypical presentations and laboratory shortcomings. Arch Intern Med. 2000;160(20):3045–53.

    Article  CAS  Google Scholar 

  51. Friedman TC. An update on the overnight dexamethasone suppression test for the diagnosis of Cushing's syndrome: limitations in patients with mild and/or episodic hypercortisolism. Exp Clin Endocrinol Diabetes. 2006;114(7):356–60.

    Article  CAS  Google Scholar 

  52. Tomlinson ES, Maggs JL, Park BK, Back DJ. Dexamethasone metabolism in vitro: species differences. J Steroid Biochem Mol Biol. 1997;62(4):345–52.

    Article  CAS  Google Scholar 

  53. Dimaraki EV, Jaffe CA. Troglitazone induces CYP3A4 activity leading to falsely abnormal dexamethasone suppression test. J Clin Endocrinol Metab. 2003;88(7):3113–6.

    Article  CAS  Google Scholar 

  54. Keitner GI, Fruzzetti AE, Miller IW, Norman WH, Brown WA. The effect of anticonvulsants on the dexamethasone suppression test. Can J Psychiatr. 1989;34(5):441–3.

    Article  CAS  Google Scholar 

  55. Chortis V, Taylor AE, Schneider P, Tomlinson JW, Hughes BA, O'Neil DM, Libé R, Allolio B, Bertagna X, Bertherat J, Beuschlein F, Fassnacht M, Karavitaki N, Mannelli M, Mantero F, Opocher G, Porfiri E, Quinkler M, Sherlock M, Terzolo M, Nightingale P, Shackleton CH, Stewart PM, Hahner S, Arlt W. Mitotane therapy in adrenocortical cancer induces CYP3A4 and inhibits 5α-reductase, explaining the need for personalized glucocorticoid and androgen replacement. J Clin Endocrinol Metab. 2013;98(1):161–71. https://doi.org/10.1210/jc.2012-2851.

    Article  CAS  PubMed  Google Scholar 

  56. Gibbons JA, de Vries M, Krauwinkel W, Ohtsu Y, Noukens J, van der Walt JS, Mol R, Mordenti J, Ouatas T. Pharmacokinetic drug interaction studies with enzalutamide. Clin Pharmacokinet. 2015;54(10):1057–69. https://doi.org/10.1007/s40262-015-0283-1.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  57. Yamashita F, Sasa Y, Yoshida S, Hisaka A, Asai Y, Kitano H, Hashida M, Suzuki H. Modeling of rifampicin-induced CYP3A4 activation dynamics for the prediction of clinical drug-drug interactions from in vitro data. PLoS One. 2013;8(9):e70330. https://doi.org/10.1371/journal.pone.0070330.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  58. Kajosaari LI, Jaakkola T, Neuvonen PJ, Backman JT. Pioglitazone, an in vitro inhibitor of CYP2C8 and CYP3A4, does not increase the plasma concentrations of the CYP2C8 and CYP3A4 substrate repaglinide. Eur J Clin Pharmacol. 2006;62(3):217–23.

    Article  CAS  Google Scholar 

  59. Nieman LK, Biller BM, Findling JW, Newell-Price J, Savage MO, Stewart PM, Montori VM. The diagnosis of Cushing's syndrome: an Endocrine Society clinical practice guideline. J Clin Endocrinol Metab. 2008;93(5):1526–40. https://doi.org/10.1210/jc.2008-0125.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  60. Qureshi AC, Bahri A, Breen LA, Barnes SC, Powrie JK, Thomas SM, Carroll PV. The influence of the route of oestrogen administration on serum levels of cortisol-binding globulin and total cortisol. Clin Endocrinol. 2007;66(5):632–5.

    Article  CAS  Google Scholar 

  61. Fritz KS, McKean AJ, Nelson JC, Wilcox RB. Analog-based free testosterone test results linked to total testosterone concentrations, not free testosterone concentrations. Clin Chem. 2008;54(3):512–6. https://doi.org/10.1373/clinchem.2007.094870.

    Article  CAS  PubMed  Google Scholar 

  62. Rosner W, Auchus RJ, Azziz R, Sluss PM, Raff H. Position statement: Utility, limitations, and pitfalls in measuring testosterone: an Endocrine Society position statement. J Clin Endocrinol Metab. 2007;92(2):405–13.

    Article  CAS  Google Scholar 

  63. Favresse J, Burlacu MC, Maiter D, Gruson D. Interferences with thyroid function immunoassays: clinical implications and detection algorithm. Endocr Rev. 2018;39(5):830–50. https://doi.org/10.1210/er.2018-00119.

    Article  PubMed  Google Scholar 

  64. Barkan A, Koenig RJ. Clinical case seminar: unraveling the mystery of abnormal thyroid function tests. Clin Diabetes Endocrinol 2015;1:14. doi: https://doi.org/10.1186/s40842-015-0010-8. eCollection 2015.

Download references

Acknowledgements

The figures in this manuscript were prepared using pre-made icons from BioRender (www. biorender.com).

Funding

None.

Author information

Authors and Affiliations

Authors

Contributions

RAH and ALB reviewed the literature and prepared the manuscript. RAH prepared the Figs. DG reviewed and edited the manuscript. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Ariel L. Barkan.

Ethics declarations

Ethics approval and consent to participate

Not applicable.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Haddad, R.A., Giacherio, D. & Barkan, A.L. Interpretation of common endocrine laboratory tests: technical pitfalls, their mechanisms and practical considerations. Clin Diabetes Endocrinol 5, 12 (2019). https://doi.org/10.1186/s40842-019-0086-7

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s40842-019-0086-7

Keywords